Teledyne LXT-220 User Manual
Page 27
Dissolved Oxygen Controller
Operation
Teledyne Analytical Instruments
27
1 DO
.00 ppm
Cal
.0 mV
 
 
Zero Calibration Menu DO
3.2.4 Parameter Selection Menu
This menu allows you to select what function you would like to
perform.
With the cursor on BUFFER, if you press ◄ and ► together the
LXT-220 will enter the calibration menus (standardize and span).
Scrolling to and choosing SETUP brings you to where you can
configure and setup the instrument. The STATUS section allows the 
user to view the calibration engineering data (mV input from the sensor 
and the zero potential from the last cal). Each section is entered by 
placing the cursor under the first letter of the section you wish to enter 
and pressing the two horizontal calibrate keys simultaneously. The 
Parameter Selection Menu is discussed in detail in Section 3.4.2. 
3.2.4.1 C
ALIBRATION
S
UBMENU
Choosing BUFFER from the Parameter Selection Menu brings up a
submenu that allows the user to perform a 1 or 2 point buffer calibration 
(standardize & span). This menu is described in detail in Section 3.4.2 
Calibration. 
Buffer 
Setup 
Status 
 
Parameter Selection Menu
 
 
 
 
Graph