4. Immediately before making the calibration adjustment, note the
current value on the display and add the differential to the 
current value. This "total adjusted" value is entered as the buffer 
value. 
3.4.2.5
C
ALIBRATION
E
RROR
D
ETECTION
& E
LECTRODE
O
PERATIONAL
G
UIDELINES
If the slope calibration performed in Section 3.4.2.3 is not within
the predetermined limits of the instrument, a flashing asterisk (*) will 
appear. 
The presence of the asterisk indicates a potential calibration
problem. In general, if the asterisk appears, the integrity of the 
electrode, the buffer solutions or the handling procedure should be 
questioned. See the dissolved oxygen sensor manual for specifics on the 
electrode. 
3.4.2.6 B
ACK
-
TO
-F
ACTORY
C
ALIBRATION
This feature allows the LXT-220 to be reset to "typical electrode"
for electrode potentials. As the electrode ages it will deteriorate and the 
mV/FS will decay and will eventually require replacement. If the 
membrane is torn or coated the calibration will be either sluggish or out 
of range. 
The following procedure for returning the LXT-220 to factory
calibration can be used for both standardize and span calibrations.
Procedure at the Standardization Menu:
1. Position the cursor under the "C" in "Cal." 
2. Press both horizontal CALIBRATE keys simultaneously and 
observe the cursor briefly move to the right. The default setting 
is complete.