Troubleshooting – Teledyne SnackO2 - Food package analyzer User Manual

Page 16

Advertising
background image

Teledyne Analytical Instruments

16

Troubleshooting

Problem Solutions

No display.

•

Check the position of
the Power switch.

•

Check battery life.

Pump is very slow or will
not run.

•

Check battery life

•

Check for clogged or
blocked sample line.

•

Check for clogged or
blocked needle.

•

Check for dirty filter.

•

Check for clogged
hose inside analyzer
and internal filter.

•

Adjust pump speed.

Unstable sensor reading.

•

Check for clogged or
blocked sample line.

•

Ensure Sample
exhaust port is not
blocked.

Sample pump stops
immediately after starting
it.

•

Increase Sample
Duration time in the
Parameter Menu.

•

Check for dirty filter or
blocked sample line.