A. Function Keys
= [AutoTare] = Enter
B. Calibration Procedure
1. Remove jumper JP3 and place on only one pin to calibrate the scale.
Displays:
Definition:
2. Press and hold [UNITS], [ZERO], or [AutoTare] while
powering ON. The system goes to Calibration mode.
CAL
3. Press the [AUTOTARE] key to select the AUTO ZERO mode
AO 0
None
Press the [UNITS] key to select the desired choice.
AO 1 AZT
0.6d
AO 2 AZT
1.0d
AO 3 AZT
2.0d
4. Press the [AUTOTARE] key to step to the WEIGHING UNITS
Un i t.LB
lbs
mode. Press the [UNITS] key to select weighing units in
Un i t.kG
kg
lb or kg mode.
5. Press the [AUTOTARE] key to step to the CAPACITY mode.
C XXXXX
Press the [UNITS] key to select the appropriate Capacity.
Section 4: Calibration