Fairbanks Series III Bench Scales User Manual

Page 9

Advertising
background image

50778

9

07/10 -- Rev. 6

A. Function Keys

= [AutoTare] = Enter

B. Calibration Procedure

1. Remove jumper JP3 and place on only one pin to calibrate the scale.

Displays:

Definition:

2. Press and hold [UNITS], [ZERO], or [AutoTare] while

powering ON. The system goes to Calibration mode.

CAL

3. Press the [AUTOTARE] key to select the AUTO ZERO mode

AO 0

None

Press the [UNITS] key to select the desired choice.

AO 1 AZT

0.6d

AO 2 AZT

1.0d

AO 3 AZT

2.0d

4. Press the [AUTOTARE] key to step to the WEIGHING UNITS

Un i t.LB

lbs

mode. Press the [UNITS] key to select weighing units in

Un i t.kG

kg

lb or kg mode.

5. Press the [AUTOTARE] key to step to the CAPACITY mode.

C XXXXX

Press the [UNITS] key to select the appropriate Capacity.

Section 4: Calibration