Cable Wire Color
Function
Balance Junction Box
Black
(-) Excitation
- E
Green
(+) Excitation
+ E
Yellow
Shield
SH
White
(+) Signal
+ S
Red
(-) Signal
- S
g.
Connect the main interface cable as follows:
Cable Wire Color
Function
Balance Junction Box
Black
(-) Excitation
- EXC
Green
(+) Excitation
+ EXC
Yellow
Shield
SHLD
White
(+) Signal
+ SIG
Red
(-) Signal
- SIG
*
(-) Sense
- SEN
*
(+) Sense
+ SEN
* = Used only when main interface cables exceed twenty five (25) feet.
h.
Tighten all gland bushing nuts
i.
Calibrate as necessary. See Platform Cornering Adjustments.
j.
Replace the box cover and tighten all screws. Replace the platform
access cover .
2.
Platform Cornering Adjustments
a.
Adjust the analog indicator to the platform. First, ensure all corners are
within one (1) division of each other at 25% of rated capacity. Follow the
appropriate indicator service manual for indicator calibration.
1.)
Perform a course platform calibration close to the actual weight.
2.)
Identify the platform corner numbers.