Appendix 1: data output specifications – Fairbanks Ultegra Health Scale User Manual
Page 8

Appendix 1:
Data Output Specifications
This appendix contains detailed data output information on the Ultegra.
RS232-C Serial Data Output
A.
Introduction
The single RS232-C output port is designed for interface with customer’s data
processing equipment.
B.
Specifications
• RS232-C compatible Data Signal
• 7 data bits
• Odd Parity
• 9600 Baud Rate (+0.1%)
• Two Stop Bits
• US-ASCII Character Set
• Mark = -3 to -15V
• Space = + 3 to +15V
• Maximum distance of 50 cable-feet
• Connector (at end of the attached cable)
DB9 Female for connection to a PC serial port
C.
Character Frame
Characters are transmitted in an ASCII format at 9600 baud (+ 0.1%). The receiver
must be capable of a tolerance of 9600 (+/- 2.5%) baud to allow for line losses and
frequency skew. Character frame consists of one start bit, 7-bit character length, odd
parity bit and two stop bits.
08/08
8
50735 Rev. 6
3 to 15 VDC
ON State (Spacing)
Logic State
-3 to -15 VDC
OFF State (Marking)
S
T
A
R
T
B
I
T
0
ON
OFF OFF
D
0
L
S
B
D
1
D
2
D
3
D
4
D
5
D
6
P
A
R
I
T
Y
B
I
T
S
T
O
P
B
I
T
0
S
T
O
P
B
I
T
0