Fault Codes & Troubleshooting

Fault Codes
If the DiscovR unit displays a fault light, the code is determined

by counting the number of times the fault light flashes.

Code 1 = Calibration unstable

Code 3 = Calibration error

Code 4 = Temperature over the range

Code 5 = Calibration compensation error

If any of these codes display, disconnect the unit from its power

source and bring it to a room temperature (60°– 80° F) place.

Check the vehicle battery voltage, and confirm it is within

specified range. Fully squeeze the rubber calibration bulb ten

(10) times, and allow the unit to cool for thirty (30) minutes.

Reconnect the unit to the refrigerant source and retest.
If the fault light persists, or if Code 6 is displayed, contact the

Technical Services Department.

Troubleshooting
no Power
– Check for correct battery clip polarity, and full

connection of the plug connector into the unit.
Excess Air/Fail – Check the sample hose for good connec-

tions to the refrigerant source and the unit. Also, check the

brass filter for any oil residue, and replace it if necessary.

Note: Clean the coupler with compressed air before installing a

new filter.