Principle of Operation

SKC Re-usable PPI Samplers are impaction-based
fi lter samplers that perform precise size-selection
for either thoracic or respirable dust, depending
on the model. PPI Samplers contain four small
impactors in the inlet section of the device. Each
impactor features a unique 50% cut-point to target
a specifi c one-quarter segment of the ISO/CEN
curve resulting in a precise fi t along the entire
curve. A sample pump operating at 2 or 8 L/min
(2 L/min only for thoracic) pulls air through the
inlet nozzle of each impactor in the inlet plate.
Particles larger than each impactor’s 50% cut-
point are scrubbed and retained by impaction
onto the porous oiled impaction substrate
contained in each impactor. Smaller particles
continue to the standard 37-mm collection fi lter
for analysis. See www.skcinc.com/prod/225-380.
asp for performance graphs.

Media Preparation

Filters: Condition and weigh fi lters according to method used. Record the weight
as the pre-sample weight.

Impaction Substrate: Using an oiled impaction substrate reduces particle
bounce. Replacement pre-oiled disposable porous plastic discs are available as SKC Cat.
No. 225-388 (pk/200).

Inserting a Collection Filter into the PPI

The PPI will arrive already assembled. Disassemble it to
insert collection fi lter.

1. Unscrew and remove two screws that hold the inlet

plate to the base plate.

2. Lift off inlet plate to expose impaction plate.

3. Lift off impaction plate to expose base plate.

Inlet plate

Impaction
substrates (4)

Standard 37-mm
collection fi lter
and support pad

Base plate with
exhaust to pump

Dust enters here