Pxd series digitizer – Teledyne LeCroy PXD Series User Manual
Page 38

34
ISSUED: March 2003
901217-01
PXD Series Digitizer
Table 2. Four Channel Models
Table 3. Typical Input Current (Amps) for Different Modules
Bandwidth Limiter: 20 MHz and 200 MHz
Sensitivity: 40 mV to 8 V full scale range
Scale Factors – volts (calibrated): 0.04, 0.08, 0.16, 0.4, 0.8, 1.6, 4, 8 FSR
Full Scale Range: 8 major divisions
Model
PDX514
PDX214
PDX114
Bandwidth
500 MHz
250 MHz
150 MHz
Maximum Single Shot Sample
Rate
1 GS/s
1 GS/s
1 GS/s
Maximum Repetitive Sample
Rate
50 GS/s
50 GS/s
50 GS/s
Channels
4
4
4
3U PXI Slots
3
3
3
Acquisition Memory Standard
256k
256k
256k
Acquisition Memory Option 1
4M
4M
4M
Single Shot Capture Window
10 ns –
10,000 s
10 ns –
10,000 s
10 ns –
10,000 s
Repetitive Capture Window
5 ns – 10 µs
5 ns – 10 µs
5 ns – 10 µs
Sequential Mode Max
Segments
4096
4096
4096
Power Consumption Maximum
70 W
70 W
70 W
PXI Chassis
Voltage (V)
PXD114/214/
514 (A)
PXD522/1022
(A)
PXD212/512
(A)
PXD1021 (A)
5
7.3
6.6
4.2
3.86
3.3
5.8
4.4
3.7
3.0
12
0.28
0.14
0.11
0.08
-12
0.35
0.25
0.22
0.17