3. Touch the Histogram button at the bottom of the dialog. After
touching this button, the Math selection for Histogram pop-up is
shown. Select the desired Math Trace in which to store the results.
Touch the Close button when you are finished.
Touch the trace descriptor label for the math trace just created for the
results, and then touch the Histogram tab to access its right-hand dialog.
Numbers on the previous image correspond with the following
explanations.
1. On the Buffer side of the Histogram right-hand dialog, touch
#Values and provide a value.
2. On the Scaling side of the Histogram right-hand dialog, touch #Bins
and provide a value.
3. Center the histogram by touching the Find Center and Width
button.
4. Instead of using the Find Center and Width button, you can provide
specific Center and Width values using their respective controls and
providing values using the pop-up numeric keypad.