Operator’s manual, Step description intermediate data test result – Teledyne LeCroy AP034 User Manual
Page 39

Operator’s Manual
922261-00 Rev A
33
Model:
AP034
Serial Number:
Asset or Tracking Number:
Date:
Technician:
MODEL
SERIAL NUMBER
CALIBRATION
DUE DATE
OSCILLOSCOPE
DIGITAL MULTIMETER
LEVELED SINE WAVE GENERATOR
HF SINE WAVE GENERATOR
FUNCTION GENERATOR
1
N/A
1
The function generator provides stimulus for making relative measurements. The output amplitude of the generator is
measured with the DMM or oscilloscope in the procedure. Thus, the generator is not required to be calibrated.
Step
Description
Intermediate Data
Test Result
Gain Accuracy
A-7
Sine Wave Generator Output Voltage
________________ mV
A-13
Probe Output Voltage
________________ mV
A-15
Gain Error
(Test Limit ≥ ±2%)
_________________ %
High Frequency Common Mode Rejection Ratio (CMRR)
B-13
Common Mode Signal at 100 MHz
________________ mV
B-15
CMRR at 100 MHz (Test Limit ≥ 18:1)
_________________ :1
Low Frequency Common Mode Rejection Ratio
C-7
Common Mode Signal at 70 Hz
________________ mV
C-8
CMRR at 70 Hz (Test Limit
≥ 10 000:1)
_________________ :1
C-14
Common Mode Signal at 1 MHz
________________ mV
C-15
CMRR at 1 MHz (Test Limit
≥ 100:1)
_________________ :1