Procedure, A. adjust coarse dc balance (r226) – Teledyne LeCroy AP034 User Manual

Page 44

Advertising
background image

AP034 Active Differential Probe

38

922261-00 Rev A

Procedure

A. Adjust Coarse DC Balance (R226)

1. If necessary, set the probe offset to 0.000 V by rotating the O

FFSET

knob in the CHANNEL

section of the oscilloscope.

2. Insert the two-pin shorting plug into the socket on the Logic Board. Use the two holes

closest to the probe cable (Figure 9).

Figure 8, AP033 Logic Board

3. To enter the Cal Mode, push either of the two offset buttons that protrude through holes in

the Logic Board.

4. Plug the precision 50

Ω

terminator into the DMM input. Set the DMM to measure DC volts.

5. Using a low capacitance screwdriver, adjust Coarse DC Bal (R226) on the Amplifier Board

for an output voltage of 0 mV ±10 mV (Figure 10).

Figure 9, Amplifier Board Adjustment Locations