Teledyne LeCroy DA1855A User Manual

Page 48

Advertising
background image

DA1855A Differential Amplifier

42

922258-00 Rev A

When viewed at 500 m/div, the same 400 V square wave shows the probe compensation to be
slightly overpeaked.

Figure 5

When the time/div is decreased to value normally used to view 20 to 150 kHz switchmode power
conversion circuits, the slightly peaked LF compensation appears as a DC level shift.

Figure 6

Viewing a power FET’s saturation voltage with the slightly peaked LF compensation makes the
voltage appear to go negative. In this example the repetition rate of the power supply is 60 kHz.

Advertising