Operator’s manual – Teledyne LeCroy DA1855A User Manual
Page 85

Operator’s Manual
922258-00 Rev A
79
High Frequency CMRR
Step
Description
Intermediate data
Test Result
5-e
Amplifier Output Voltage at 10 MHz
_____________ V
5-f
Differential Mode Gain at
10 MHz
___________
5-w
Common Mode Feedthrough at 10 MHz
___________
5-y
Common Mode Gain at
10 MHz
___________
5-aa
Common Mode Rejection Ratio at 10 MHz (Test limit:
≥ 1000:1)
__________:1
Low Frequency CMRR
6-m
Common Mode Feedthrough at 70 Hz
_____________ V
6-n
Common Mode Gain at
70 Hz
___________
6-o
Common Mode Rejection Ratio at 70 Hz (Test limit:
≥ 50,000:1)
__________:1
6-x
Common Mode Feedthrough at 100 kHz
_____________ V
6-y
Common Mode Gain at
100 kHz
_____________ V
6-z
Common Mode Rejection Ratio at 100 KHz (Test limit
≥ 50,000:1
__________:1
Precision Voltage Generator Accuracy
7-c
PVG Zero Output Voltage
(Test limit: 0.05% of reading + 500 μV)
__________ V
7-g
PVG Output Voltage at +15.5 V
(Test limit: 0.05% of reading + 500 μV)
__________ V
7-j
PVG Output Voltage at
–15.5 V (Test limit: 0.05% of reading + 500 μV)
__________ V