ADP30x Probe
Performance Verification
This procedure can be used to verify the warranted characteristics of the
ADP30x series Active Differential Probes. 
If the product does not meet specifications, it should be returned to a Teledyne 
LeCroy service center. There are no user accessible adjustments, so there is no
adjustment procedure.
Test Equipment Required
The following table lists the test equipment and accessories (or their
equivalents) that are required for performance verification of the ADP30x Active
Differential probe. This procedure is designed to minimize the number of
calibrated test instruments required. Because the input and output connector
types may vary on different brands and models of test instruments, additional
adapters or cables may be required. 
Only the parameters listed in boldface in the "Minimum requirements" column 
must be calibrated to the accuracy indicated.
Description
Minimum Requirements
Test Equipment Examples
Digital Multimeter
DC: 0.1% Accuracy 
AC: 0.2% accuracy to 
measure 700 mV at 70 Hz 
Agilent Technologies 
34401A 
Fluke 8842A-09 
Keithley 2001 
Function Generator
Sine Wave output waveforms 
20 Vp-p 
Agilent Technologies 
33120A 
Stanford Research Model 
DS340 
Leader LAG-120B 
BNC Coaxial Cable
Male-to-Male 50
Ω Cable
Pomona 5697-36
Calibration Fixture
ProBus Extender Cable
Teledyne LeCroy PROBUS-
CF01 
Banana Plug Adapter
Female BNC-to-Dual Banana 
Plug 
Pomona 1269 or Mueller 
BU-00260 
Insulated Banana Plug 
Insulated Banana 
Couplers (2) 
Insulated BNC-to-Shrouded 
Banana Plug 
Mueller BU-5671-B-12-0 
Mueller BU-32601-2 (Red) 
Mueller BU-32601-6 (Blue) 
22