Proper compensation of the probe is required to assure good amplitude
accuracy in the dynamic portions of the waveform being measured. LF
compensation matches the probe to differences in oscilloscope input
capacitance. The LF compensation should always be checked and adjusted as
needed when first connecting a passive probe to the oscilloscope input. HF
compensation matches time constants within the probe to compensate for
normal component tolerances. It is typically not necessary to adjust HF
compensation unless the probe is being used with an oscilloscope with large
differences in input characteristics than the oscilloscope model it was designed
for.

LF compensation is performed by connecting the input of the probe to a low
frequency square wave, such as the oscilloscope calibrator signal set to 1 kHz.
The compensation is adjusted by rotating the adjustment accessible through the
small hole in the center of the housing near the BNC connector. Use the tool
supplied with the probe for this adjustment.

Undershoot

Overshoot

Correct adjustment

Should HF compensation be required, access the
adjustments by sliding the black plastic cover off
the compensation housing near the BNC con-
nector. A pulse generator with low overshoot and
a 300 ps risetime is the required signal source,
along with a set of attenuators. The probe must
be connected to a terminated probe tip to BNC
adapter.

Some overshoot and ring will be present at some
settings of V/Div.Adjust both trimmers for the overall best response on all
ranges.

Typical optimum HF

adjustment