Tip Selection and Exchange

The PP009 probe is supplied with a rigid tip. The spring tips PK1-5MM-109 (0.8
mm) or PK1-5MM-124 (0.38 mm) combine sharp points with on-axis compliance
and can be ordered directly from Teledyne LeCroy as optional accessories. All
three tip styles provide reliable connections under a wide range of physical
interconnect situations. While lacking the on-axis compliance feature of the
spring tips, the rigid tip has a larger diameter and is more robust when exposed
to physical stress at the tip. Select the proper tip for your application needs.

To change the tip or replace it when damaged, carefully grip the outer most
portion of tip and pull straight out, along the axis of the probe, using needle
nose pliers. Do not attempt to grip the plastic insulator with pliers when
removing the tip, as this will squeeze the tip, which will make it difficult or
impossible to remove. Do not grip the outer gold plated tube which the tip slides
into. With the tip removed, align the replacement tip with the hole and begin
the insertion with the pliers. The tip may be fully seated by placing the probe
against a hard surface and gently applying pressure.

To remove tip, grip the spring-loaded

portion of the tip beyond the outer sleeve

and pull straight out.

Do not apply pliers to the plastic insulator.