QPHY-ENET Software Option

QPHY-ENET Operator’s Manual Rev G

49

Figure 37. Block diagram of TF-ENET-B fixture section D

1. Calibrate data path (see the

X

Signal Path Calibration Procedure

X

topic on page

X

74

X

).

2. Calibrate disturber sine wave (see the

X

Disturber Calibration Procedure

X

topic on page

X

77

X

).

3. Connect AWG to J3 and J4.

4. Connect DUT to J68.

5. Install jumpers on J17/J19, J99/J100, J91/J92 and select Pair for your test.

Table 5. Section D test pair jumper configurations

6. Transmit signal with disturbing sine wave should look like the picture shown in

X

Figure 41

X

. Select

1000Base-T under “Select Standard.”

7. Select Mode 1 Transmit Wform in the “Select Test” field.

8. Check Differential Data on 2 Channels box and enter C2 for Source for + Data and C3 for Source for

– Data.

9. Touch the Set Up and Start Test button to begin testing. The test automatically attempts to position each

pulse, so that it passes its template.

Pair

Install

Remove

A

J39, J40

J61, J60, J62

B

J43, J44

J59, J60, J62

C

J41, J42

J59, J61, J62

D

J45, J46

J59, J61, J60