Write protection – Yokogawa BT200 User Manual

Page 3

Advertising
background image

Write Protection

Brain Communication

Field Guide

FEBRUARY 2006

‘NO’. This indicates that this transmitter is
NOT password protected.

Section 2.4

Recovery Password


When you lose or miss-place your password,
it is possible to release the Password
Function by using the general password

YOKOGAWA

”. When this password is

used, you have 10 minutes to reset the
password only. No other parameters can be
changed during this time. Reset your
password per the procedure in

Section 2.1

.

Section 3.0

Software protection of the

external Zero Adjust Screw.

To disable the external Zero Adjust Screw,
set the parameter J15: EXT ZERO ADJ to
INHIBIT.

Note: If the Password Feature is enabled, the
password must be entered before the
parameter J15 can be changed.

Section 4.0

Transmitters with Displays


EJX Transmitters with LCD display will
indicate when the Write Protect is enabled.
A “Key” symbol will appear on the display
when the Write Protect is enabled.

(See Figure 4)

Section 5.0

Setting When Shipped

Hardware Write Protection (WR Switch):

E” (Write Enabled)

Software Write Protection (Password

Feature):

De-activated

External Zero Adjust Screw:

Enabled








A “Key” symbol will
appear when the Write
Protect is enabled.

Figure 4