background image

IM 11Y01B02-11E-A 3rd Edition :August 14, 2012-00

3-4

<3 INSTALLATION AND WIRING>

6

Factory

1

Ethernet

+ Transmit

2

- Transmit

3

+ Receive

4

- Receive

7

Factory

1-8

Detect

Internal connections only – do not use

8

Field

Analog

#1 output

+

Cal/Val initiate signal loop (SV # 1) to Remote voltage free con-
tacts/switch. Do not apply external power!

-

Analog

#2 output

+

Cal/Val initiate signal loop (SV # 2) to Remote voltage free con-
tacts/switch. Do not apply external power!

-

Analog

#3 output

+

Cal/Val initiate signal loop (SV # 2) to Remote voltage free con-
tacts/switch. Do not apply external power!

-

9

Factory
and/or
Field

1

Gas

Temperature

Compensation

Externally powered 4-20 mA gas temperature signals are wired to
1 (+) and 2(-), 3 not used.
Loop powered temperature transmitter can be connected to 1 (-)
and 3 (+24 VDC), 2 not used

2

3

4

Gas

Pressure

Compensation

Externally powered 4-20 mA gas pressure signals are wired to 1
(+) and 2(-), 3 not used.
Loop powered pressure transmitter can be connected to 1 (-) and
3 (+24 VDC), 2 not used

5

6

14

Factory

1-4

Mini-Display

Internal connections only – do not use

 

Figure 2 – Internal Wiring Diagram, including optional cell heating and optional purge unit