IM 11M6D2-02E-A

10.2.2 Pressure Regulator, M1132KD

This general purpose regulator is used to adjust the pressure of instrument air entering the AC4/AC8. Figure 10.4

Figure 10.4 Pressure Regulator, M1132KD

Flow Capacity:

20 SCFM (33.6m_/hr) at 100 psig supply – 20 psig outlet

Exhaust Capacity:

0.1 SCFM (0.17m_/hr) – downstream pressure 5 psig above set point

Sensitivity:

1” of water

Effect of Supply Pressure Variation: Less than 0.2 psig for 25 psi change
Maximum Supply Pressure:

250 psig

Air Consumption:

Less than 6 SCFM

Output Range:

0 to 60 psi

Port Size:

1/4” NPT

Materials:

Body – Die cast aluminum alloy; Diaphragm – Nitrile elastomer

and nylon fabric; Trim – Brass, zinc plated steel, acetyl

Weight:

4.0 lbs (1.8 kg)

10.2.3 Pressure Gauge, M1132ME

The purpose of this gauge is to indicate the pressure of instrument air entering into the AC4/AC8. Figure 10.5

Figure 10.5 Reference Air Gauge, M1132ME

Gauge Size:

2” (50mm)

Measuring Range:

0 to 60 psi

Connection:

1/4” FNPT

Weight:

0.5 lbs (0.2 kg)