1.3 PIPING

Parts list:

2 M1132RR ss male connection fittings
1 IAC-24 –22-E-A calibration unit
1 Instruction sheet

1.4 Mounting the IAC-24

1) Wrap the threads of the M1132RR with Teflon tape and thread into the ZR22G calibration and reference ports.
2) Push the male connectors into the IAC-24 device (IAC24/ZR22G assembly should resemble Fig.1)

Tighten the fittings on the IAC-24 to seal the IAC-24 onto the male connections.

3) Connect the 1/4” tube to the IAC-24. Instrument air must be 21% air 02.
4) Set the inlet gas pressure to the IAC-24 to 7 PSIG +/-2 PSIG.

Fig. 1

Important Note: To prevent leakage all threaded fittings should have teflon tape (or suitable alternate).

1.5 Wiring
CAUTION:
The IAC-24 requires 24VDC. It will fail if 120VAC is used for power. Consult the factory for
directions to convert from an AC to a DC autocalibration system.

The IAC-24 requires a special electrical connector. This is purchased separately as an integral part of
the cable. For a new installation of the ZR402G and ZR22G system the WZ-H-6H-[ ] heater cable should
be used with the analyzer/detector combination. This cable has the wire and connections for both the
detector heater and the 24VDC autocalibration accessory. This cable may be used with both the
ZR402G single point analyzer and the AV550
G multipoint analyzer.