Wiring – Yokogawa FU20/PH20 All-in-One pH/ORP Sensor User Manual

Page 16

Advertising
background image

IM 12B6J3-E-H

12

TEMPERATURE

TEMPERATURE

REFERENCE

REFERENCE SCREEN

SOLUTION GROUND

Pt (ORP Measure)

GLASS SCREEN

(11)

(12)

(13)

(17)

(14)

(15)

(16)

16

15

11

12

13

14

(Not used)

T2

T1

RE

SE

GE

S

ORP WIRING DIAGRAM with normal reference

TEMPERATURE

TEMPERATURE

REFERENCE (pH)

REFERENCE SCREEN

(SOLUTION GROUND)

Pt (ORP Measure)

GLASS SCREEN

(11)

(12)

(13)

(17)

(14)

(15)

(16)

13

11

12

15

14

(Not used)

16

T2

T1

RE

SE

GE

S

ORP WIRING DIAGRAM with pH sensor as reference

TEMPERATURE

TEMPERATURE

REFERENCE

REFERENCE SCREEN

SOLUTION GROUND

GLASS (Measure)

GLASS SCREEN

(11)

(12)

(13)

(17)

(14)

(15)

(16)

11

12

13

14

16

T2

T1

RE

SE

GE

S

15

pH (& ORP) WIRING DIAGRAM

6. WIRING

6-1. Conventional pH (& ORP) Wiring
Connect the PH20 or FU20 to the EXA or
EXAxt PH Analyzer as shown (Fig 12). With
this configuration, it is possible to measure
ORP (or rH) at the same time (Refer to the
EXA or EXAxt Manual for appropriate impe-
dance jumper and Service Code settings).

6-2. Wiring for ORP measurement with

normal reference

Connect the PH20 or FU20 to the EXA or
EXAxt PH Analyzer as shown (Fig 13).
Refer to the EXA or EXAxt Manual for
appropriate impedance jumper and Service
Code settings).

6-3. Wiring for ORP measurement with

pH reference

Connect the PH20 or FU20 to the EXA or
EXAxt PH Analyzer as shown (Fig 14).
Refer to the EXA or EXAxt Manual for
appropriate impedance jumper and Service
Code settings).

Figure 12.

Figure 13.

Figure 14.

Advertising