Installation, 1 connections of vp single coax cable – Yokogawa SC24V Differential pH ORP Electrode User Manual

Page 8

Advertising
background image

IM 12B6W2-02E-E

8

3. Installation

3.1 Connections of VP single coax cable

The numbers in the table refer to the numbers of the terminals in the instrument.

A B C D E F Jumper Instrument

FU24/FU20: pH+ORP(rH) 15

13 16 14

11

12

FU24/FU20: pH / ref

15

13 16 14

11

12

PH202 / PH402

FU24/FU20: ORP / ref

-

13 16 15

11

12

13-14 PH450 / PH150

FU24/FU20: ORP / pH

13

- 16 15

11

12

15-14

Polagraphic DO sensor

(OXYgold)

17

18

-

14

11

12

-

DO202

Galvanic DO sensor

13

15

- (14) 11

12

-

DO402

2-electrode

conductivity

13-14

sensor

15

16

13

14

11

12

15-16 SC402 / SC202

4-electrode conductivity

SC450 / SC150

sensor

16

15

13

14

11

12

-

Tabel 2

The numbers (11- 17) in the table refer to the numbers on the terminals in EXA Instruments.
Other instruments are indicated in the following tables

PH100

function

pin ID

Color

name

number

pH

A

Transparent GE

12

Ref

B

Brown

RE

14

C

Black

Ground el.

D

Yellow

SE

16

TC

E

Red

T1

17

TC

F

Blue

T2

18

Tabel 3

A

B

D (yellow)

E (red)

F (blue)

C (black)

(transparant)

(brown)

Cable

A

B

C D E

F

brown

yellow

black

red

blue

transparant

Connector

Transparant

Brown

Black

Yellow

Red

Blue

Figure 1