Glue – Yokogawa FS20 Fittings for pH and ORP User Manual

Page 12

Advertising
background image

IM 12B6K1-01E-E

3-3-2. Mounting the subassembly
The subassembly can be cemented or welded directly in a piping system or in a T-piece of
it. Install the subassembly at a convenient location for maintenance and calibration.
Ensure that there is some place at the top of the fitting (approx. 20 cm) for mounting or
replacing electrodes.
Fig. 3-6 shows some mounting examples. Select the mounting position so that the sensors
are immersed in the process liquid during measurement.

(Model WU20) should be used (see fig. 3-7).

Figure 3-6

3-4. Sensor mounting
3-4-1. General
For stable measurement the glass- and the reference electrode should be mounted in the
holes nearest to the earth connection.

When mounting sensors of Yokogawa in a fitting or subassembly the electrode cables
(Model WU20) should be used (see fig. 3-7).

Figure 3-7

10 cm

colour marker

screen blue

core red

10 cm

colour marker

colour marker

outer screening white

core red

inner screening blue

Cable nut

Cable nut

colour marker

COAX-cable (model WU20-PC) for single electrode

TRIAX-cable (model WU20-LT) for combined electrode

GLUE

GLUE

GLUE

PVC

10

3-hole (DN50)

4-hole (DN80)