2 assignment of power supply terminals, Calibration, 1 items to check before power on – Yokogawa JUXTA VJCE User Manual

Page 10: 2 connection of calibration equipment, Caution, Te-16 terminal block

Advertising
background image

10

All Rights Reserved. Copyright © 1999, Yokogawa M&C Corporation

IM 77J01C51-01E

2nd Edition July 23,2004-00

5.2

Assignment of Power Supply Terminals

1

1

2
3

2

3

Terminal Number

Signal Symbol

SUPPLY

L (

ϩ

)

SUPPLY

N(

Ϫ

)

GND

CAUTION

Ensure that the power supply voltage for VJCE-01A matches that for the signal conditioner to be
mounted on VJCE-01A. Supply of different power supply voltage may damage VJCE-01A and signal
conditioners.

6. CALIBRATION

Refer to the User’s Manual of respective signal conditioner for how to calibrate and for the equipment required for

calibration.

6.1

Items to Check before Power on

Supply power rating is 12 to 36 V DC or 85 to 264 V AC / DC.

Wiring of signal cables

Installation, ambient temperature, humidity, dust, vibration

Please power on after checking the above items.

The VJCE would be in operational status upon power on. However, 10 to 15 minutes are required to satisfy its speci-

fications and performance

6.2

Connection of Calibration Equipment

When Output- or Output-2 is connector, prepare KS2 cable and TE-16 terminal block.

KS2 cable

43

Cable with 16 pairs shield

t=15

80

[VJCE side]

40 pin connector

40 pin connector

[Input/output card side]

Ground terminal

TE-16 Terminal Block

1A

2A

3A

4A

5A

6A

7A

8A

1B

2B

3B

4B

5B

6B

7B

8B

9B

10B

11B

12B

13B

14B

15B

16B SHLD

9A

10A

11A

12A

13A

14A

15A

16A SHLD