4. SETTING INPUT FILTER

The input filter is used to suppress the influence of chattering

noise or the like.

Setting: Set the filter switch on the front panel of the signal

conditioner to “ON.” (The factory default setting is “OFF”)

• The input filter, time constant (approx.10ms), is connected.

However, the input frequency is limited to 100 Hz or less, and

the pulse width is 3ms or more.

“OFF” setting

“ON” setting

5. MAINTENANCE

The product starts running immediately when the power is turned

on; however, it needs 10 to 15 minutes of warm-up before it

meets the specified performance.

5.1 Calibration Apparatus

• A calibrator (Yokogawa CA150 or equivalent)

5.2 Calibration Procedure

1. Connect the instruments as shown below. First check the

output-1 signal and then the output-2 signal.

10

11

3 2 1

4

5

6

7

8

9

Power supply

11

10

8

L

+

N

GND

3

4

+

Input

Calibrator

Calibrator

2

5

Output-2

+

Calibrator

7

9

Output-1

+

2. Produce a rectangular pulse of any frequency from the

calibrator to measure the value using a calibrator, then check

that the frequency is output.