3. EXTERNAL WIRING

WARNING

Be sure to turn OFF the power supply before wiring to

avoid the risk of electric shock. Use a tester or similar

device to ensure that no power is being supplied to a

cable to be connected.

M4 screw terminals are provided for the connection of external
signals. Attach a crimp-on lug to each wire for connection to the
terminals.

• Recommended cables: A nominal cross-sectional area of

0.5 mm

2

or thicker for signal cables, and that of 1.25 mm

2

or

thicker for power cables.

Wiring Diagram

1

2

4

5

6

7

6
7

4
5

+

Output

Input

Power Supply

+

–−

1
2

PS+

+

NOTE

Do not connect anything to the terminals that are

not used in the wiring diagram. Otherwise it may

cause the malfunction or damage.

The power line and input/output signal lines should

be installed away from noise-generating sources.

Otherwise accuracy cannot be guaranteed.

Adhere strictly to the specifications to avoid

overheating or damage. Before turning on the

power, ensure the following:

Power supply voltage and input signal value

applied to the product should meet the required

specifications.

The external wiring to the terminals and wiring to

ground are as specifications.

Do not operate the product in the presence of

flammable or explosive gases or vapors.

This product is sensitive to static electricity;

exercise care in handling. Before you operate the

product, touch a nearby metal part to discharge

static electricity.

Power Supply and Isolation

Power supply voltage: 24 V DC±10% (percentage ripple: less

than 5% p-p)

Current consumption : 65 mA
Insulation resistance: 100 MΩ at 500 V DC between (input and

output) and power supply.

Withstand voltage:

500 V AC/min. between (input and output)

and power supply

4. MAINTENANCE

The product starts running immediately when the power is turned

on; however, it needs 10 to 15 minutes of warm-up before it

meets the specified performance.

4.1 Calibration Apparatus

• A voltage current source (Yokogawa GS200 or equivalent)

• A digital multimeter, DMM (Yokogawa 7561 or equivalent)

• A precision resistor of 250 Ω±0.01%, 1W

4.2 Calibration Procedure

1. Connect the instruments as shown below.

1

2

4

5

6

7

2

4
5

+

+

Ro

DMM

L+ N–

Power

supply

24V DC

Voltage

current

source

Ro: a precision resistor 250Ω

Input

Output

Ro is used for current output (FA4A).

2. Use the voltage current source and apply the electromotive

force equivalent to 0, 25, 50, 75, and 100% of the measuring

range to the distributor.

3. Verify that the corresponding output voltages are 0, 25, 50,

75, and 100% respectively and within the specified accuracy

rating.