INTRODUCTION

This manual describes the function, operation and maintenance of the Nitrogen Tester. The
intent of this manual is to describe the function of the Nitrogen Tester only. The final assembly
manufacturer should provide operating instructions for the completed assembly. The Nitrogen
Tester is engineered for long life, maximum reliability and stable performance.

NOTE: In order to obtain optimum performance from the Nitrogen Tester, all operation and

maintenance must be performed in accordance with this manual. Please read the
manual thoroughly before using the Nitrogen Tester and do not attempt any repair
or procedure that is not described herein. The warranty does not cover any damage
resulting from misuse, unauthorized repair or improper maintenance of the Nitrogen
Tester.

WARNING:

Never allow an excess length of tubing or any accessory near a person’s head
or neck, which may result in strangulation. Before use, all individuals who will
be using the Nitrogen Tester must become thoroughly familiar with the
information contained in this manual. Strict adherence to the operating
instructions is necessary for safe, effective product performance. This product
will perform only as designed if installed and operated in accordance with the
manufacturer’s operating instructions.

Use only genuine accessories and replacement parts. Failure to do so may
seriously impair the performance. Repair or alteration of the Nitrogen Tester
beyond the scope of the maintenance instructions, or by anyone other than an
authorized service person, could cause the product to fail to perform as
designed.

Calibrate the Nitrogen Tester weekly, when in operation, or if environmental
conditions change significantly (Elevation, Temperature, Pressure, Humidity —
refer to “Factors Influencing Accurate Readings”).

WARNING:

Use of the Nitrogen Tester near devices that generate electrical fields may
cause erratic readings

WARNING:

DO NOT over pressurize the sensor. Doing so may destroy the sensor and
void the warranty. To avoid over pressurization only allow 5 psi (or 2 liters per
minute) of gas to come in contact with the sensor membrane.

WARNING:

Ensure proper tire inflation pressure after use, if required.

WARNING:

The oxygen sensor is a sealed device containing a mild acid electrolyte, lead
(Pb), and lead acetate. Lead and lead acetate are hazardous waste
constituents and should be disposed of properly, or returned for proper
disposal or recovery.

WARNING:

Do not immerse the device in any cleaning solution, autoclave or expose the
sensor to high temperatures greater than 150

0

F (70

0

C).

WARNING:

Dropping the device can adversely affect its performance.