Page 5

CAL Err lo: Sensor voltage too low

Press and hold the Calibration Button for three seconds to manually force a new calibration.
If unit repeats this error more than three times, contact RTI for possible sensor replacement.

CAL Err hi: Sensor voltage too high

Press and hold the Calibration Button for three seconds to manually force a new calibration.
If unit repeats this error more than three times, contact Customer Service for possible sensor
replacement.

CLEANING AND MAINTENANCE

Store the Nitrogen Tester in a temperature similar to its ambient environment of daily use.

The instructions given below describe the methods to clean the instrument, sensor and its
accessories:

Instrument

! When cleaning or disinfecting the exterior of the Nitrogen Tester, take appropriate care

to prevent any solution from entering the instrument. Do not immerse unit in fluids.

Oxygen Sensor

! Clean the sensor with a cloth moistened with a 65% alcohol / water solution.

! It’s not recommended to use spray disinfectants because they can contain salt, which

can accumulate in the sensor membrane and impair readings.

Accessory

! The threaded barbed adapter may be cleaned by washing it with a 65 percent alcohol

& water solution. The part must be thoroughly dry before it is re-used.

SPECIFICATIONS

Measurement Range: 0-99.9%

Nitrogen

Resolution: 0.1%

Nitrogen

Accuracy and Linearity:

1% of full scale at constant temperature, R.H. and pressure when
calibrated at full scale.

Total Accuracy:

±3% actual nitrogen level over full operating temperature range.

Response Time:

90% of final value in approximately 15 seconds at 23

/C.

Warm-up Time:

None required

Operating Temperature:

15

/C - 40/C (59/F - 104/F)

Storage Temperature:

-15

/C - 50/C (5/F - 122/F)

Humidity

0-95% (non-condensing)

Sensor Type:

Galvanic fuel cell

Battery Type:

3V Lithium button-cell (Non-replaceable)

Expected

Sensor

Life:

>900,000 O2 percent hours minimum

2-years

in

typical

applications

Weight:

Approximately 3 ounces