UMC 600IS Installation Manual

1.2

System Limitations and Restrictions

The following items represent limitations and restrictions on the use of the UMC 600IS:

•

All instrumentation must have identification
nameplates in compliance with the control
drawing.

•

All wiring, connections, conduit and grounds
must comply with the National Electrical
Code.

•

The cable type and cable length must comply
with the specifications in the control drawing

•

No modifications can be made in the field.

•

Factory Mutual Approval required (see
Section 1.1 on page 1).

•

It is mandatory to return the UMC 600IS to
Condec for service.

•

Component level repair is not permitted on
Factory Mutual Approved equipment.

•

Standard unit configuration is without local
sense jumpers installed. Therefore, use either
six-conductor loadcell cable or externally
jump +SEN to +EXC and –SEN to –EXC.

1.3

Front Panel Keypad and Annunciators

Figure 1-1 shows the UMC 600IS front panel. The UMC 600IS display consist of a six digit, seven-segment
display. Table 1-1 lists the front panel keys and their functions.

Figure 1-1. UMC 600IS Front Panel

1

6

2

7

3

8

4

9

5

0

CE

ENT

TARE

RECALL

ZERO

NET

GROSS

TARE

lb/kg

CONV

PRINT

1 SET

POINT

2 SET

POINT

CAPACITY

600 x 0.1

ZERO

NET

GROSS

lb

kg

MOTION

Panel Key

Function

ZERO

Provides push-button auto zero (PAZ) over ±1.9% or 100% full scale capacity. Operates only in gross
weighing mode.

NET/GROSS

Switches the unit between gross and net weighing modes.

TARE

Provides push-button tare entry over 100% of scale capacity. Pressing

TARE

key switches to net mode

and enters tare.

PRINT

Provides a manual print function if unit is wired to serial printer or other data device.

lb/kg CONV

Switches the displayed weight unit between pounds and kilograms.

TARE RECALL

Press to recall tare value; LED annunciator flashes when tare value is displayed.

ON/OFF

Provides power to the indicator.

SETPOINT 1 & 2

No function. Setpoints are not supported for this indicator.

0–9, CE, ENT

Numeric keyboard for entry of manual tare and calibration data. Includes CE (Clear Entry) and ENT (Enter)
keys.

Table 1-1. Front Panel Key Functions