2 technical data – LumaSense Technologies M390 User Manual

Page 10

Advertising
background image

M390 Manual

Introduction 10

2.2 Technical Data

This

Model

Model

Part

No. Exit

Aperture Temperature

Range Power

[

] M390-A1 14029-A1 0.625

in

600-2300 °C

15KVA

[

] M390-A2 14029-A2 1

in

600-2300 °C

15KVA

[

] M390-B1 14029-B1 0.625

in

600-2600 °C

15KVA

[

] M390-B2 14029-B2 1

in

600-2600 °C

15KVA

[

] M390-C1 14029-C1 0.625

in

600-3000 °C

15KVA

[

] M390-C2 14029-C2 1

in

600-3000 °C

24KVA

[

] M390-L1 14029-L1 1

in

300-2000 °C

15KVA

Accuracy:

±0.25% of reading ±1 C

Cavity:

Closed end graphite tube 150 mm (6.0") long with 64 mm (2.50")
heated length

Source Diameter:

16 mm (.625"), 25 mm (1.00")

Emissivity
(effective from 0.6 to 1.8 µm)

1.00, ~0.96 from 3 to 15 µm

Cooling:

Water cooled. Standard hose connection on rear, 3 lpm (1 gpm)

Cooling Inlet Pressure:

90 PSI maximum

Customer is responsible to provide pressure safety valves at inlet
and outlet. Outlet (drain) must not be blocked.

Purge Gas:

Argon High Purity (99.99 % pure)

Purge Gas Inlet Pressure:

25 PSI maximum. Customer is responsible to provide over
pressure safety valves at inlet

Heating Element Type:

Graphite tube with argon gas purge, 80 lph (4-8 cfh)

Temperature Sensor:

Mikron M668L (or M680 for M390L1)

Method of Control:

Digital PID Controller

Remote Set Point:

RS232 serial communication output (RS485 optional)

Warm-up Time:

5 minutes from ambient to 2300 °C

Operating Ambient Temp.:

0 C to 44 C (32 F to 110 F)

Operating Ambient
Humidity:

0 to 90% RH non-condensing

Power Requirements:

[ ] 208 or [ ] 220 or [ ] 240 VAC ±10% [ ]50 [ ]60Hz

[ ] 15 kVA [ ] 24 kVA

Power Connector:

Customer supplied.
[ ] 15kVA

Must use plug that complies with IEC-309 at 75 Amps at 220VAC

[ ] 24kVA

Must use plug that complies with IEC-309 at125 Amps at 220VAC

Dimensions:

171 cm H x 56 cm W x 82 cm D

Maximum Weight:

182 kgs (400 lbs)

Advertising