3 installation, 1 installing the m390, 2 water and purge gas supply requirements – LumaSense Technologies M390 User Manual

Page 11: 1 water, 2 purge gas

Advertising
background image

M390 Manual

Installation 11

3 Installation

3.1 Installing the M390

To install the M390:

1. Allow the M390 to warm-up to room temperature for at least 2 hours after unpacking

and prior to applying power to remove any condensation.

2. Connect the power cable to a VAC service that matches the power requirement

specification noted on the SPECIFICATION LABEL located on the rear of the M390.

3. Connect the inlet water hoses to a 80 psi city or other clean water supply. The Customer

is responsible to provide pressure safety relief valves at inlet and outlet.

4. Connect the outlet water hose to an open drain. Outlet (drain) must not be blocked.
5. Connect the gas purge to a pre-purified Argon gas tank. The Customer is responsible to

provide pressure safety relief valves at inlet.

6. Set the regulator to 15 psi.
7. Turn on the water and check for leaks.
8. Check all connecting liquid, gas and electrical connections for integrity.

Note: The Blackbody Cavity (or tubular heating element) may have been shipped
separately. Please use included instructional video files in the enclosed CD to properly
install the cavity assembly.

Note: If the element or quartz tube is damaged or requires replacement, please see

the maintenance section for cavity replacement procedure.

3.2 Water and Purge Gas Supply Requirements

3.2.1 Water

Flow 1 GPM (3.8 l/m), Pressure: 90 PSI (6.3 kg/cm2) maximum
Customer is responsible to provide pressure safety valves at inlet and outlet. Outlet (drain) must
not be blocked.

3.2.2 Purge Gas

High Purity (99.99 % pure) Argon. (See Warning Note Below). Customer is responsible to provide
pressure safety valves at inlet.

Warning: DO NOT USE NITROGEN (N

2

) FOR PURGE GAS. AT HIGH TEMPERATURES,

CYNIDE GAS COULD BE GENERATED.

Pressure: Regulated at 15 PSI (1.8 kg/cm2)