12.5.1 Retaining Existing Calibration Factors
When zero-point, humidity-interference and cross-interference
calibrations are performed, all the installed filters are calibrated. In
some case, you may wish to keep the calibration factors already
present in your monitor. If this is the case, use the following
instructions.
Fig.12.8 Schematic diagram showing the calibration task Set-Up