4 configuring the monitor, 1 setting the clock – LumaSense Technologies INNOVA 1314i User Manual
Page 64

Chapter 5
______________________________________________________________________
_____________________________________________________________________________
BE6030-13
1314i Photoacoustic Gas Monitor
LumaSense Technologies A/S
Page 64 of 199
S1 By pressing this push-button the user can come out of reset
mode without performing any kind of reset. The Monitor just
goes back to operating the same way it was operating before
RESET was pressed.
S3 By pressing this push-button the user confirms that a FULL
reset is required. During a FULL reset the Monitor performs the
following two tasks:
·
“Clearing” (emptying) all data from its Working Memory.
This means all data in Display Memory and Background
Memory will be lost.
·
Copying data from its Source Memory into its Working
Memory. This means that set-up parameters in block 3 of
Working Memory “Other Set-up Parameters” will be given
their default values, and the values of the optical filter set-
up parameters in Working Memory will be the same as
those found in Source Memory.
5.4 Configuring the Monitor
5.4.1
Setting the Clock
Before dispatch from the factory, the internal clock of the Monitor is
set to the correct local (Danish) time.
The current date is entered as Year-Month-Day for example the
24
th
of July 2014 is entered as 2014-07-24.
The current time is entered as the time on the 24-hour clock for
example 14:05 is five minutes after two in the afternoon. Table
5.2 details the range of acceptable values and the default value of
the date and time.
PERFORM FULL RESET ?
NO YES