SANDPIPER RuppTech Stroke Counter/Batch Control User Manual

Page 3

Advertising
background image

Batch Control/Stroke Counter

Page 3

Batch-sc-rev0611

520-180-000 3/02

Batch Control / Stroke Counter

Page 3

Programming for Accuracy

To preset accurate batch programs, it is important to know actual system

performance. Gather this information by measuring output volumes gener-
ated by test runs of the systems, using published pump performance as
a starting program. Careful measurement will tell you how much volume
your system is actually discharging on every stroke. You can then adjust
the program accordingly. Here is an example.

EXAMPLE OF USE:

A product is blended based on incoming orders. Two ingredients are

mixed in a ratio of 3 to 1, creating the final product.

Order volumes are typically 50, 500, 1000, 1500 and 2000 gallons. Two

pumps transfer the ingredients to a mixing area for “batching”.

Pump #1 in this example is the transfer pump for the main element. Its

control has Preset Counts set as follows:

Pump #1 Preset Counts:......... PO - 37.5 gallons ..... P1 - 375 gallons
................................................ P2 - 750 gallons ....... P3 - 1125 gallons.
................................................ P4 - 1500 gallons
Stroke Scale Valve: ................. 0.34 gallons per stroke

The control’s Stroke Scale Value is set at the displacement per stroke of
the 1½-inch pump to which it is connected, .34 gallons.

Pump #2 in this example is the transfer pump for the second element.
Its control has Preset Counts set as follows, based on the 3 to 1 mixing
ratio:

Pump #2 Preset Counts:......... PO - 12.5 gallons ..... P1 - 125 gallons
................................................ P2 - 250 gallons ....... P3 - 375 gallons
................................................ P4 - 500 gallons
Stroke Scale Valve: ................. 0.09 gallons per stroke

Pump #2’s control has the Stroke Scale Value set at the displacement

per stroke of the 1 -inch pump to which it is connected, .09 gallons.

Though each pump requires its own control unit, both controls can be

tied together with one remote Start/Stop switch running both units.

TESTING THE SYSTEM

Since each control gets one pulse or signal

from its pump for each stroke, the volumes
referenced in the example should be set for the
closest total volume available.

For example:
Pump #1.............P1 - 375 gallons @
............................0.34 gal. per stroke
............................= 1102.94 strokes

Adjusting for 1103 strokes would deliver

375.02 gallons, assuming 100% efficiency from
pump and system. Realistic expectations of a
well-designed and implemented system fall in
the range of 1% to 3%.

Based on data obtained from the system,

it becomes evident where the system will
perform from an accuracy and efficiency
standpoint. This will indicate adjustments to
be made.

For example:
Testing showed in the example system

that Pump #1 delivered 363.93 gallons, within
the 3% range of accuracy. Since the Stroke
Scale Value was set at .34 gallons per stroke,
1103 strokes actually produced .329 gallons
per stroke.

Formula is:
Flow ÷ Strokes = Gallons per Stroke

363.93 ÷ 1103 = 0.329 gal./stroke

This test indicates the system should be

reset to produce a total flow closer to the batch
target of 375 gallons.

Formula is:
Batch Target ÷ Gal. per Stroke = # Strokes

375 ÷ 0.329 = 1139.81

Rounded to 1140 strokes

The controller cannot count a fractional

stroke, so the number of strokes is rounded
to the nearest whole number, 1140. This gets
closer to the target batch size of 375 gallons,
actually producing 375.06 gallons.

The same type of testing, calculation and

adjustment would be applied to all pumps in
a system. Each pump’s performance will be
effected by the application and system
in total.

WHERE TO FIND PUBLISHED PUMP

PERFORMANCE

Published pump performance is found on the data sheet
issued for model pump. Displacement is shown to the right
of the performance curve.

If you do not have the appropriate data sheet, contact your
local Warren Rupp or Marathon distributor for a copy. Or
you can contact the Factory Literature Department at the
number shown below.

To receive the correct sheet, be sure to verify the pump
model and type, which is shown on the pump nameplate.

Advertising