Model 2200p ph analyzer / controller, Technical data, Related products - probes & accessories – Aquametrix 2200P pH Analyzer/Controller Datasheet User Manual

Page 2: Ordering information

Advertising
background image

Model 2200P pH Analyzer / Controller

www.Aquametrix.com

978-749-9949

Toll free - 855-747-7623 (855-pHProbe)

Technical Data

Related Products - Probes & Accessories

Display

4 x 7 segment 1/2” LED Display

Indicator Lights

4 LED indicators: Status, Alarm, Relay A,

Relay B

Measuring Range

pH

0.00 to 14.00 pH

Temperature

0 to 40

0

C (32 to 104

0

F)

Power Requirements

Standard

98-132 VAC 50/60 Hz, (less than 10 VA)

Optional

196-264 VAC 50/60 Hz (less than 10 VA)

23-26 VDC (nominal current: 150 mA)

Ambient Conditions

Temperature

-30 to 50°C (-22 to 122°F)

Humidity

0 to 90% RH non-condensing

Control Relays

Number

Two Control Relays: Relay A, Relay B

Rating

5 A 115/230 VAC, 5A 30 VDC, SPDT

Bidirectional

Relays factory preset for bidirectional

control, can be individually field set for ris-

ing or falling process

On/ Off Setpoints

Field selectable full scale

Fail-safe

Normal or fail safe operation

Alarm Relay

Number

One Alarm Relay

Rating

5A 115/230 VAC, 5A 30 VDC, SPDT

Alarm High

Field selectable 0 -14 pH

Alarm Low

Field selectable 0 -14 pH

Deadband

Fixed at 2% of full scale

Fail Safe

Normal or fail safe operation

Relay Indicators

Three individual LEDs indicate status

of two control relays and alarm relay

Analog Outputs

Non-isolated

0-1 mA, 100 ohms maximum load

Non-Isolated

0-5 VDC, 1000 ohms minimum load

Isolated

4-20 mA, 800 ohms maximum load

Output is isolated from the input, the

ground, line power and all other outputs.

Range Expand

Analog outputs can be scaled to rep-

resent any segment of the instrument

scale, minimum of 10% of full scale.

Output Hold

Analog outputs are placed on hold

when the instrument is in menu mode.

Temperature

The 0-1 mA and 0-5 VDC outputs can

be selected to track either the pH or

the temperature of the process.

Sensor to Controller Distance

Differential

P60 Series Differential pH Probe: maximum

distance 3000 feet (914 m)

Combination

500 Series: maximum distance 25 feet (7.62 m)

Calibration

2 point, manual or automatic buffer recognition

Temperature Compensation

Automatic

0-100°C (32 to 212°F) if probe has temperature

sensor.

Diagnostics

Invalid Entries

Identified by flashing LEDs

System Error

Two color STATUS LED indicates system sta-

tus. Green no fault, Red system fault. Scroll to

Status in Menu to determine fault code. Refer

to Manual for probable cause and remedy.

Alarm Relay

Alarm relay can be enabled to activate upon

high and low process values and memory loss.

Test

Display value and analog outputs can be set to

any value for testing and diagnostic purposes.

Safety & Security

Operator

Password protected (activated by DIP switch)

Memory

Nonvolatile memory (EPROM)

Microprocessor

Watchdog timer monitors microprocessor. In-

strument automatically returns to online opera-

tion if left in menu mode for more than 10 min-

utes and no key is pressed.

Sensitivity

0.1% of span

Stability

0.1% of span per 24 hours non-cumulative

Temperature Drift

Zero - 0.01% of span per °C

Span - 0.01% of span per °C

Enclosure

NEMA 4X molded fiberglass reinforced polyes-

ter enclosure with four 1/2” conduit holes and

mounting feet for surface mount. A NEMA 4X

plug is provided for one hole.

Mounting

Standard

Surface Mount

Optional

Panel Mount kit P/N C35-68

Pipe Mount kit P/N C35-69

Weight

3.5 lbs. (1.6 kg)

P60C-6

Industrial quality differential pH probe with replace-

able salt bridge, temperature sensor and integral

preamplifier. Compression fitting 1-1/4” NPT.

P60C-8

Industrial quality differential pH probe with replace-

able salt bridge, temperature sensor and inte-

gral preamplifier. Fixed threads 1-1/2” NPT.

P575

Industrial quality combination style pH

probe. Fixed threads 1” NPT.

P60-HTC

CPVC Wet/Hot tap ball valve assembly

STC60

Tank mount hardware for pH/ORP differential probes

2200P

pH Analyzer/Controller

1

120 VAC, 50/60 Hz Power

2

240 VAC, 50/60 Hz Power

A

Surface Mount

B

Panel Mount

C

Pipe Mount

2200P

Ordering Information

Advertising