Model p/r60c-4 differential ph or orp probes – Aquametrix P/R60C-4 Differential pH or ORP Probes User Manual

Page 2

Advertising
background image

Model P/R60C-4 Differential pH or ORP Probes

www.WaterAnalytics.net

978-749-9949

Toll free - 855-747-7623 (855-pHProbe)

Technical Data

Related Products

Measuring Range

pH

0 to 14 pH

ORP

-1000 to +1000 mV

Flow Rate

10 ft./sec maximum (3 metres/sec)

Flow should be as low as possible in low con-

ductivity water and in solutions with high sus-

pended solids

Wetted Materials

CPVC, kynar, glass, titanium palladium alloy

and EPDM (platinum for ORP probe)

Transmission

Distance

3000 ft. (900 m)

Sensitivity

pH

0.01 pH

ORP

0.1 mV

Stability

pH

0.03 pH per day, non-cumulative

ORP

2 mV per day, non-cumulative

Temperature Compensation

Automatic

-5 to 95°C (23 to 203°F)

Pressure Limit

100 psig at 65°C maximum

Temperature Limits

CPVC

-5 to 95°C (23 to 203°F)

The temperature limit of probes in flow-through

applications is limited by pressure and by the

pipe fitting material.

Probe Cable

5 Conductor plus shield, 15 ft. (4.6 m) long

CABLES & ACCESORIES

JB-1

NEMA 4X junction box

C42-5PXXX Interconnect cable; specify length

C35-17K

Salt bridge kit w/ kynar outer junction

(Package of 3)

C35-17

Salt bridge kit w/ ceramic outer junction

(Package of 3)

CALIBRATION SOLUTIONS

A35-13

pH 4 Buffer, 500 mL.

A35-14

pH 7 Buffer, 500 mL

A35-24

pH 10 Buffer, 500 mL

A35-40

ORP Buffer, 200 mV, 500 mL

A35-41

ORP Buffer, 600 mV, 500 mL

Ordering Information

P60C-4

pH probe, 1 1/2” NPT threaded body

P60C-4-H

pH probe with Hard glass electrode

P60C-4-A

pH probe with Antimony electrode

R60C-4

ORP probe, 1 1/2” NPT threaded body

R60C-4-H

ORP probe with Hard glass electrode

R60C-4-G

ORP probe with Gold electrode

Differential Measurement Technique

Dimensions

Advertising