BEFORE YOU BEGIN

Button Functions

!

READ BUTTON - Reads the Percent Moisture

Content value (%MC), corrected for temperature
and species.

@

CALIBRATION CHECK BUTTON - Checks

meter calibration. It also displays the average
of up to 100 accumulated readings; displays the
maximum stored reading; erases the readings.

#

SPECIES BUTTON - Sets the species code

for the wood you are using. Species are numbered
from 1 to 48 and are listed on the Species
Code Chart. This button also acts as a scroll
key, depending on the function.

$

TEMPERATURE BUTTON - Sets the wood

temperature and changes the temperature
mode (Fahrenheit or Celsius). This button also
acts as a scroll key, depending on the function.

CHECK CALIBRATION

Press the calibration check button

@

and read button

!

simultaneously. Meter is in calibration if it displays 12%
(+ or - .2).

If you check the calibration and the meter does not display
12% it is likely an indication of a low battery. If this occurs,
change the battery immediately. Continued use with a low
battery may cause the meter to go out of calibration. If you
have a fresh battery and the instrument still does not indicate
a proper calibration, return it to DELMHORST for service.
See

Service for your Meter section.

When the battery is removed and then reconnected, the
meter displays its software version for one second and then
turns itself off. After replacing the battery, you must reset
the meter as described in

Resetting the Meter section.