7. Based on the application and use of relay contact outputs, complete all wiring terminations prior to application

of power. Shut-in controls may be omitted until system test is complete. Terminals are labeled “ALARM 1”,
“ALARM 2”, “ALARM 3”, and “FAULT” (COM & NO/NC). Relay contact outputs may be discreet or zoned
via the gold-plated jumper tabs.

8. If applicable, terminate a remote mounted normally open momentary switch to the terminations labeled “Alarm

Reset”. Activation of the alarm reset function will reset the alarms of all Model 12B controllers. The reset func-
tion is effective when the Model 12B’s respective alarms have been programmed in the latching position and
alarm conditions have passed. Each Model 12B controller has its own alarm reset switch which is discussed fur-
ther in section 2.0.