Trigger, Input, Enable – GW Instek GLA-1000 Series Quick start guide User Manual
Page 19: Data
Advertising

Trigger
GLA-1016
GLA-1032
GLA-1132
Channel
16
32
32
Condition
Edge/Pattern Edge/Pattern Edge/Pattern
Pre/Post
trigger
0% ~ 100% 0% ~ 100% 0% ~ 100%
Level
1
1
1
Threshold
+6V ~ −6V +6V ~ −6V +6V ~ −6V
Accuracy
±93mV
±93mV
±93mV
Count
1 ~ 65535
1 ~ 65535
1 ~ 65535
Page
Maximum
8191
Maximum
8191
Maximum
8191
Input
GLA-1016
GLA-1032
GLA-1132
Maximum
Input Voltage
±30V
±30V
±30V
Impedance
500kΩ/10pF 500kΩ/10pF 500kΩ/10pF
Enable
GLA-1016
GLA-1032
GLA-1132
Channel
16
32
32
Enable
Condition
Don’t care,
Low, High
Don’t care,
Low, High
Don’t care,
Low, High
Enable Delay
1 ~ 65535ms 1 ~ 65535ms 1 ~ 65535ms
Data
GLA-1016
GLA-1032
GLA-1132
Skew
<1.5ns
<1.5ns
<1.5ns
Compression
Channels
16
24
24
Compression
Ratio
Maximum
255
Maximum
255
Maximum
255