GOS-6103C OSCILLOSCOPE
USER MANUAL
51
(g).Typical △ Θ cursor function for phase
measurement.
A phase measurement is done by first setting
a reference for the full 360
o
waveform
period:
5div=360
o
.
NOTE. When the VOLTS/DIV or the TIME/DIV controls are in
uncalibrated setting, the △V and △T measurement values will be
displayed with divisions.
When the vertical mode is set to the ADD mode, and the CH1 and CH2
VOLTS/DIV controls are set to different scales, the △V measurement
values will be displayed with divisions.
GOS-6103C OSCILLOSCOPE
USER MANUAL
52
AUTO MEASUREMENT
Proceed a measurement by using the auto-measurement according to the
following steps:
1. Pressing and holding the MEAS’MT FUNCTION pushbutton to turn on the
auto-measurement and measurement readout.
2. Briefly pressing the pushbutton to select the four measurement functions in
the sequence as below:
FREQ—PERIOD—WIDTH—DUTY—OFF
And refer to the following figure: