54
Tracking Generator (Optional)
Frequency Range
100 kHz to 3 GHz
Output Power
-50 dBm to 0 dBm in 0.5 dB steps
Absolute Accuracy ± 0.5 dB
@160 MHz, -10 dBm,
Source attenuation 10 dB,
20 to 30°C
Output Flatness
Referenced to 160 MHz, -10 dBm
100 kHz to 2 GHz
± 1.5 dB
2 GHz to 3 GHz
± 2 dB
Output Level
Switching
Uncertainty
± 0.8 dB
Referenced to -10 dBm
Harmonics
< -30 dBc
Typical, output level = -10
dBm
Reverse Power
+30 dBm max.
Connector type
N-type female
Impedance
50 ohm
Nominal
Output VSWR
< 1.6:1
300 kHz to 3 GHz, source
attenuation ≥ 12 dB
USB Power Sensor (Optional)
Type
Average power sensor Model: PWS-06
Interface to Meter USB cable to GSP930 Front-Panel USB Host
Connector Type
N-type male, 50 ohm nominal
Input VSWR
1.1: 1
Typical
1.3: 1
Max
Input Frequency
1 to 6200 MHz
Sensing Level
-32 to +20 dBm
Max. Input Damage
Power
≥ 27 dBm