2.9 STATus subsystem

GW APS-1102 STAT Cond.vi
This returns the operation condition register (OPCR). The response range is 0 to 20482.
<Output>

OPCR

This outputs the operation condition register (OPCR).

GW APS-1102 STAT Enab.vi
This sets the operation event enable register (OPEE). The setting range is 0 to 20482.
<Input>

OPEE set

This sets the operation event enable register (OPEE). The default setting is “0”.

<Output>

OPEE

When Read = True, queries contents of the operation event enable register (OPEE) set for the
APS-1102. When Read = False, “99999” is set.

GW APS-1102 STAT Oper.vi
This returns the operation event register (OPER) value. The response range is 0 to 20482.
<Output>

OPER

This outputs the operation event register (OPER) value.

GW APS-1102 STAT Ntr_Ptr.vi
This sets the operation transition filter (OPTF). The setting range is 0 to 20482.
 When the NTR filter is set to 1, OPER is set to 1 when OPCR changes from 1 to 0.
 When the PTR filter is set to 1, OPER is set to 1 when OPCR changes from 0 to 1.
 When the NTR and PTR filters are both set to 1, OPER is set to 1 when OPCR changes.
 When the NTR and PTR filters are both set to 0, changes in OPCR are not propagated to OPER.
<Input>

NTR set

This sets the NTR filter. The default setting is “0”.

PTR set