1gsa/s & single shot – GW Instek GDS-2000 training manual User Manual
Page 22

GDS-2000 Series Training Manual
22
1GSa/s & Single Shot
Background
By using a very fast sampling rate combined with
single shot trigger mode, you can accurately
reconstruct weak signals which are buried in
larger waveform. Here you can observe a tiny
pulse sitting on top of PECL, 3.3V DC signal.
Connection
CH1
(BNC connector on the demo board)
This demo uses a BNC to BNC cable instead of probe.
CH1 Scale
25ns/div, 50mV/div (1GSa/s)
1. Press the CH2 key twice and turn it OFF.
CH2
CH2
2. Press the CH1 key → F1 (Coupling) and select AC (~). A pulse
signal becomes visible on the display.
CH1
Coupling
F 1
Adjust the CH1 vertical position
if necessary.
POSITION
3. Press the Trigger MENU key → F3 (Mode) repeatedly and select
Single trigger mode.
Mode
Single
F 3
MENU
4. Press the Run/Stop key to initiate a single trigger. A pulse signal
becomes visible in a clear manner.