GW Instek SFG-830 User Manual User Manual
Page 41

p. 38 SFG-830
Appendix 1 Commands of IEEE488.2
*IDN? Identification Query
z After receipt of this query the instrument generates response
message as below:
z The response consists of the following four fields:
─Manufacture
─Model
─Serial Number (return 0 if not available)
─firmware Revision (return 0 if not available)
z The fields are separated by commas.
z Example:GW.Inc,SFG-830,Serial number 1111-000-90,1.0.
*RST
Reset Command
z The command performs a device reset which sets the instrument to a
defined status as bellow:
modulation OFF
frequency 1KHz
waveform sine
amplitude 1V
DC-offset OV
z The reset command shall not affect the following:
─The state of the IEEE 488.1 interface
─The selected IEEE 488.1 address of the device
─The Output Queue
─The Service Request Enable Register.
─The Standard Event Status Enable Register
─Calibration dat