GW Instek SFG-830 User Manual User Manual
Page 43

p. 40 SFG-830
*OPC? Operation Complete Query
z Place an ASCII character 1 into the instrument‘s output queue when
all pending instrument operations are finished.
*CLS
Clear Status Command
z This command clears status data structures. On the other hand, it
sets the bits of the〝Standard Event Register〞and the〝Status Byte
Register〞to zero.
z The Output Queue and the MAV bit will not be cleared.
*ESE
Standard Event Status Enable Command
z *ESE, followed by a decimal value, sets the bits of the〝Standard
Event Status Enable Register〞
which correspond to that decimal value
to 1. This enables the assigned bits of the〝Standard Event Status
Register〞.
*ESE? Standard Event Status Enable Query
z This query asks for the contents of the 〝Standard Event Status Event
Register〞.The response is a decimal value, e.g.,〝255〞means all
events of the〝Standard Event Status Register〞are enabled, in other
word, all bits are 1.
*ESR? Standard Event Status Register Query
z Ask for the contents of the〝Standard Event Status Register〞. The
response is a decimal value. The query clears the register contents.