4.3.5. Frequency

measurement


CAUTION

Maximum input for AC voltage is 750V rms. Do not attempt to take any
voltage measurement exceeding such limit to avoid the risk of electrical
shock and damages to the instrument.

Fig. 7: Frequency measurement


1. Turn the selector on Hz.
2. Insert the test leads into the jacks, the red plug into

HzV

ΩµA jack and the black plug

into COM jack (Fig. 7).

3. Connect the red and black test leads to the positive and negative poles of the circuit

under test respectively. The voltage value will be displayed with automatic detection of
the range.

4. Press RANGE key to activate the manual mode; the “AUTO” symbol disappears from

display. Press cyclically RANGE to change the position of the decimal point on the
display. To activate the Autorange function keep pressed RANGE for at least 1 second.

5. The message "OL" means that the frequency exceeds the measuring limits. In this

case disconnect the test leads from the circuit under test to avoid damaging the
instrument or endanger your own safety.

6. For minimum and maximum value measurement, HOLD and Relative

∆REL functions

please refer to paragraph 4.2.