Resistance

Range Resolution

Accuracy

Open

voltage Overload protection

400.0

Ω 0.1Ω

±(1.0%rdg+5dgt)

4.000k

Ω 0.001kΩ

40.00k

Ω 0.01kΩ

400.0k

Ω 0.1kΩ

±(0.7%rdg+2dgt)

4.000M

Ω 0.001MΩ

±(1.0%rdg+2dgt)

40.00M

Ω 0.01MΩ

±(1.5%rdg+5dgt)

1.3V 600V

rms


Diode test

Range Resolution

Accuracy

(0.4

÷ 0.8V)

Test

current

Open

voltage

Overload protection

10mV

±(1.5%rdg+5dgt)

1.5mA <3V

600V

rms


Continuity test

Range Buzzer

Overload protection

R<450

600V rms


Capacitance

Range Resolution

Accuracy

Overload protection

4.000nF 0.001nF

±(3.0%rdg+10dgt)

40.00nF 0.01nF
400.0nF 0.1nF
4.000

µF 0.001µF

40.00

µF 0.01µF

400.0

µF 0.1µF

±(2.0%rdg+8gt)

4.000mF 0.001mF
40.00mF 0.01mF

±(5.0%rdg+20dgt)

600V rms


6.1.1. Electrical

specifications

Conversion: TRMS
Measuring rate:

1.5 times per second

Temperature coefficient:

0.15×(accuracy)/°C (<18°C and >28°C)

NMRR Normal Mode Rejection Ratio:

> 50dB for DC parameters and 50/60Hz

CMRR Common Mode Rejection Ratio: >100dB from DC up to 60Hz on DCV

> 60dB from DC up to 60Hz on ACV


6.1.2. Safety
The instrument complies with:

EN 61010-1

Insulation:

Class 2, double insulation

Pollution degree:

2

Overvoltage category:

CAT IV 600V, CAT III 1000V (V/

/µA)

CAT IV 500V (A)

Max height:

2000m

Advertising