Modbus Registers

Common Registers

Register

Size

Mode &

Access Level

(R/W)

Data Type

Description

9001

1

R/W

ushort

Device ID = 31 for RDO PRO-X

9002

2

R/W

ulong

Device serial number

9004

3

R/W

time

Manufacture date

Sensor Status Registers

Register

Size

Mode &

Access Level

(R/W)

Data Type

Description

0005

3

R1

time

Cap start time
0 = no cap

0008

3

R1

time

Replace cap
0 = no cap
8 = Service period exceeded or moderate abrasion
detected on cap

Sensor health diagnostics indicate when the RDO sensor has
been damaged in the field. If the sensor has sustained moderate
damage, the probe provides a D.O value that includes a warning.
The same warning is included with readings taken after the sensor
has reached its 24-month recommended lifespan.

However, if the sensor has been severely damaged, an error
message is shown and a D.O. value is not provided. This prevents
you from receiving an erroneous reading.